RE: Right-left inversion
From: Angus Bond \(E-mail\) (afb8@alumni.cwru.edu)
Hello Elena,
Have you considered that your medical images may be using a different coordinate system than the volumizer bricks?
In standard axial (transverse) medical images, the images are presented as viewed from the feet. The pixel data is arranged in left-to-right rows (patient right-to-left), with the rows arranged top-to-bottom (patient anterior-to-posterior). If your slices are arranged bottom-to-top (patient inferior aspect to patient superior aspect), you will have a right-handed coordinate system. If your slices are arranged top-to-bottom (patient superior aspect to patient inferior aspect), you will have a left-handed coordinate system.
If the handedness of your data is the opposite of the handedness of the volumizer bricks, it will appear as if the data was swapped left-to-right.
Good luck!